Understanding qPCR

Understanding qPCR

[Part 3]

Real-time PCR analysis:

  • Baseline

The baseline of real-time PCR refers to the signal level during the initial PCR cycles, usually cycles 3 to 15, in which there is little change in fluorescent signal. The low-level signal of the baseline can be equated to the background or the “noise” of the reaction. The baseline in real-time PCR is determined empirically for each reaction by user or automated analysis of the amplification plot. The baseline should be set carefully to allow accurate determination of the threshold cycle (Ct), defined below. The baseline determination should take into account enough cycles to eliminate the background found in the early cycles of amplification, but should not include the cycles in which the amplification signal begins to rise above background. When comparing different real-time PCR experiments, the baseline should be defined in the same way for each

  • Threshold

?The threshold of real-time PCR is the level of signal that reflects a statistically significant increase over the calculated baseline signal. It is set to distinguish relevant amplification signal from the background. Usually, real-time PCR instrument software automatically sets the threshold at 10 times the standard deviation of the fluorescence value of the baseline. However, the position of the threshold can be set at any point in the exponential phase of PCR.

  • Threshold cycle (Ct)

Ct?is the cycle number at which the fluorescent signal of the reaction crosses the threshold. The Ct?is used to calculate the initial DNA copy number, because the Ct?value is inversely related to the starting amount of target. For example, in comparing real-time PCR results from samples containing different amounts of target, a sample with twice the starting amount will yield a Ct?value one cycle earlier than a sample with half the number of copies of the target. This assumes that the PCR is operating at 100% efficiency (i.e., the amount of product doubles perfectly during each cycle) in both reactions. The acceptable range for efficiency is 90-110%.?As the template amount decreases, the cycle number at which significant amplification is seen increases. With a 10-fold dilution series, the Ct?values are ~3.3 cycles apart.

  • Standard curve

A dilution series of known template concentrations can be used to establish a standard curve for determining the initial starting amount of the target template in experimental samples or for assessing the reaction efficiency. The log of each known concentration in the dilution series (x-axis) is plotted against the Ct?value for that concentration (y-axis). From this standard curve, information about the performance of the reaction as well as various reaction parameters (including slope, y-intercept, and correlation coefficient) can be derived. The concentrations chosen for the standard curve should encompass the expected concentration range of the target in the experimental samples.

  • Correlation coefficient (R2)

The correlation coefficient is a measure of how well the data fits the standard curve. The R2?value reflects the linearity of the standard curve. Ideally, R2?= 1, although 0.999 is generally the maximum value.

  • Y-intercept

The y-intercept corresponds to the theoretical limit of detection of the reaction, or the Ct?value expected if the lowest copy number of target molecules denoted on the x-axis gave rise to statistically significant amplification. Though PCR is theoretically capable of detecting a single copy of a target, a copy number of 2–10 is commonly specified as the lowest target level that can be reliably quantified in real-time PCR applications. This limits the usefulness of the y-intercept value as a direct measure of sensitivity. However, the y-intercept value may be useful for comparing different amplification systems and targets.

  • Exponential phase

It is important to quantify real-time PCR in the early part of the exponential phase as opposed to in the later cycles or when the reaction reaches the plateau. At the beginning of the exponential phase, all reagents are still in excess and the DNA polymerase is still highly efficient. The amplification product, which is present in a low amount, will not compete with the primers’ annealing capabilities. All of these factors contribute to more accurate data

  • Slope

The slope of the log-linear phase of the amplification reaction is a measure of reaction efficiency. To obtain accurate and reproducible results, reactions should have an efficiency as close to 100% as possible, equivalent to a slope of –3.32.

  • Efficiency

A PCR efficiency of 100% corresponds to a slope of –3.32, as determined by the following equation:

  • Efficiency = 10(–1/slope) – 1

Ideally, the efficiency of PCR should be 100%, meaning the template doubles after each thermal cycle during exponential amplification. The actual efficiency can give valuable information about the reaction. Experimental factors such as the length, secondary structure, and GC content of the amplicon can influence efficiency. Other conditions that may influence efficiency are the dynamics of the reaction itself, the use of non-optimal reagent concentrations, and enzyme quality, which can result in efficiencies below 90%. The presence of PCR inhibitors in one or more of the reagents can produce efficiencies of greater than 110%. An acceptable reaction should have an efficiency between 90% and 110%, which corresponds to a slope between –3.58 and –3.10.