Trace metal analysis with solid-state electrodes – Part 1
Want to learn more? Check out our blog post on this topic.
This new series of blog posts covers a range of new sensors suitable for the determination of ?heavy metals? using voltammetric methods.
The quantification of heavy metal ions plays an important role in many applications, including environmental monitoring, waste management, research studies, or even in clinical tests. Heavy metals occur naturally, but the rise of industrialization and urbanization in the past two centuries are responsible for increased levels in our environment. These dangerous elements are released and accumulate in the soil, and in ground or surface water. They enter the food chain directly from drinking water or through bioaccumulation in plants and animals. It is for this reason that pregnant women are discouraged from eating seafood, on the basis of mercury (Hg) accumulation through the food chain.
The degree of toxicity depends on the type of metal, its biological role, and most importantly, its concentration. Increased concentrations of lead, iron, cadmium, copper, arsenic, chromium, or nickel in drinking water are most often responsible for human poisoning. To highlight the toxicity of certain heavy metals in drinking water and to protect human health, guideline values or limit values for the heavy metal concentration in drinking water have been set by international organizations as the World Health Organization (WHO) or by such authorities as the U.S. Environmental Protection Agency (EPA) or the European Commission.
Several techniques have been developed for heavy metal ion analysis in the past. Commonly used techniques include atomic absorption spectrometry (AAS), inductively coupled plasma (ICP), or fluorescence spectrometry. However, these techniques require expensive equipment combined with high maintenance costs and trained personnel. Therefore, a price-effective, straightforward and sensitive method that allows detection of metal ions in water samples is highly desired.
Stripping voltammetry is the right solution for these challenges providing a simple, rapid, and cost-effective alternative for the aforementioned techniques that is also suited for untrained personnel. In addition, detection limits in the ng/L range and the possibility to determine the trace levels of heavy metals in the field make it so interesting and valuable.
The principle of stripping voltammetry
Voltammetric determination of heavy metals consists of two steps. In the first step, the analyte is preconcentrated on the surface of the working electrode as shown using the example of anodic stripping voltammetric determination of lead (Pb) in Figure 1.
Figure 1. Anodic stripping voltammetry – deposition of lead (solution stirred).
In the subsequent stripping step (Figure 2), the analyte is released. This can be achieved by oxidation or reduction depending on the method used for the determination. This step generates the analytical signal, which has to be proportional to the deposited amount of analyte.
Figure 2. Anodic stripping voltammetry – stripping of lead (solution not stirred).
Besides anodic stripping voltammetry, cathodic stripping voltammetry or adsorptive stripping voltammetry are also possible to utilize and work in a similar manner. All of these methods have something in common: every voltammetric determination is as good as the sensor used for the measurement. Therefore, in this series of posts we want to introduce our powerful sensors and demonstrate the outstanding performance with a few typical applications.
Need for new sensors
The need for heavy metal ion determinations in the field, sensor costs, and environmental issues are the main triggers for research on new sensors in voltammetry. Non-toxic and inexpensive materials are preferred for new sensors. The properties of these materials, however, can lead to some restrictions. First is the limited number of elements that can be detected on a particular electrode material (e.g., gold, carbon or bismuth). In addition, it is difficult to determine several elements simultaneously at the same mercury-free sensor. The choice of the most suitable electrode material in combination with the optimum sensor design helps to overcome these issues.
Bismuth as an alternative electrode material
In the past, there were many attempts to find less toxic electrode materials than mercury for the determination of heavy metal ions, but none have achieved exceptional electroanalytical performance. Twenty years ago, an American researcher by the name of Joseph Wang reported a bismuth film electrode for the first time (Joseph Wang, 2000).
Figure 3. Bismuth crystal.
After this initial revolutionary report, bismuth-based electrodes prepared as in-situ and ex-situ films on solid-state electrodes such as carbon, have been growing in popularity. The broad electrochemical window and low toxicity of bismuth were key factors. In addition, bismuth is able to form alloys with quite a high number of heavy metals and it exhibits high hydrogen overpotential, similar to mercury. These properties are particularly interesting for stripping voltammetry. The hydrogen evolution is suppressed very efficiently with the consequence that noise-free measurements at negative potentials can be carried out. Bismuth electrodes based on bismuth films are a good option. However, film deposition is an additional step that is time-consuming.
New sensor in VA: the Bi drop electrode
With the Bi drop electrode, a novel solid-state electrode is now available for the determination of heavy metal ions in drinking water. A bismuth drop of approximately 2 mm diameter serves as the working electrode within the voltammetric measurement.
Figure 4. Bismuth drop electrode from Metrohm.
The electrode works without the need for polishing or film deposition—only electrochemical activation is required. This significantly shortens the entire analysis time. Once activated, series of heavy metal determinations with high repeatability in the low μg/L and even ng/L range are possible.
The Bi drop electrode allows for mercury-free monitoring of the limit values of the heavy metals cadmium, lead, nickel, cobalt, and iron in drinking water. Since the electrode does not require mechanical treatment, it is especially suitable for online applications. Another advantage of the Bi drop electrode is fact that cadmium and lead as well as nickel and cobalt can be determined simultaneously.
The sensor is cost-efficient, stable, extremely sensitive, and is able to deliver more reproducible results than other previously examined bismuth-based electrodes. To demonstrate the broad possibilities and flexibility of the Bi drop electrode, examples for anodic stripping voltammetry, adsorptive stripping voltammetry, and direct voltammetric determination are presented and discussed further in the blog post.
Key features of the Bi drop electrode
- Non-toxic, completely mercury-free alternative for trace metal determination
- Simultaneous determination of Ni and Co, as well as Cd and Pb
- Limit of detection in low μg/L and even ng/L range
- Suitable for automated and online systems