Thermocouple Response times for EN285 Compliance during Qualification
The Qualification of Porous Loads / Equipment Loads to European Standards (EN285:2015) requires fast responding temperature measurement to demonstrate good air removal. Equilibration time is measured as the time between the drain temperature and the load temperature achieving the sterilisation temperature band and must be ≤15 seconds for small chambers and ≤30 seconds for large chambers.
To record the Equilibration Time accurately requires a fast data logger and fast responding temperature measurement. EN285 is quite specific in this regard with detailed specification for logging interval (1 second) and the response time of temperature measurement probes used.
Bill Bingley our Instrumentation, Calibration, Controls SME recently completed a study for a client where they had been challenged by an auditor for this information. Some of that information is shared here with the permission of the client.
The key compliance requirements from EN285 for the temperature measurement are as follows:-
The maximum cross sectional area ≤ 3.1mm2 requires a probe diameter of less than 2mm which is achievable.
The Temperature Response time is much harder to achieve. Test Method:-
Since all of the industry benchmark data loggers used for qualification log at 1 second intervals, the decision was taken to use such data loggers for this testing. Our Kaye AVS system was used for the testing.
A range of thermocouples were tested, all Type T class 1 accuracy:-
The test was performed using a stirred water bath with temperature stability at 90C +/- 0.1C. The probes stabilised at room temperature 21C before insertion. Since 1 second logging interval is used the fast acting thermocouples can be seen to have a near step change with the 1 second accuracy. However the encapsulated tipped thermocouples and the larger 4mm Stainless Steel sheath thermocouple have measurably slower response times and significantly away from the EN285 requirement. Many runs were performed to demonstrate repeatability and the results were very consistent with response times as follows:-
The conclusion; The tip encapsulation with some polymer material is significant insulation and therefore impact on response time. The large 4mm stainless steel probe is actually faster than the thermocouple with encapsulated tip. When measuring a short equilibration time this is a significant impact, the site conclusion and feedback to auditor in this instance is that encapsulated thermocouple will not be used for Porous Load Qualification.
It is often argued that provided all the thermocouples have an equally slow response then this doesn't matter? But it is certainly a long way from EN285 compliance and also a potential variable in the critical measurement that should be eliminated if possible.
Validation Engineer @ Bristol-Myers Squibb | Equipment Qualification
3 年Most loggers when using a full sim will not let you log at 1 second intervals. Running at 3/5 second intervals will mean enough data points will be reported with potential to show issues within equilibration time. If there is an issue, air trapped somewhere, you will see it at 3/5 second intervals, why is 1 second interval required? Also, a delay in tc response time should still show any failures within the 30 seconds required time. Tipped TCs are more robust. Welded or non tipped tcs often open after long use time and cause issues.