Temperature Sensor Calibration

Temperature Sensor Calibration

Calibration is an integral part of thermal validation and should be optimised for the process and sensors being used.

To ensure optimal calibration of our sensors we use a variety of calibration baths, RTDs and Bath Inserts to ensure the optimal performance of the sensors.

When working with highly accurate sensors, the calibration bath is critically important to ensuring a good calibration. Some of our sensors have accuracies of ± 0.05°C and we perform testing from -196 to 400°C, so choosing the correct temperature bath is critical for effective calibration of our sensors.

In addition to temperature ranges and accuracy we must also consider the design of the thermal wells within the calibration block. Oversized wells allow air between the sensor and the well insulating the sensor and reducing the accuracy of the calibration. This is often seen when doing low temperature calibrations and frost builds up in the airspace in the well insulating the sensor. For Example, calibrating 2mm thermocouples in a 4mm thermal well will result in a poor-quality calibration. The sensors must therefore be placed in a suitable calibration well.

Jofra RTC-159 Calibration Inserts

For Thermocouple calibration we typically use a Kaye 140 or 400 Calibration Bath. The accuracy of a thermocouple is around ±0.5°C so these baths are ideal for this. The wells of the Kaye 140 and 400 bath are also suitable for the size of most Thermocouples.

Calibration Inserts for Ellab and Vaisala Sensors

For the calibration verification of our Ellab sensors we have a 24 port Jofra Insert drilled with an Insulation cover. This provides extremely accurate and repeatable results from -100 to 150°C, with typical temperature deviations of around ±0.01-0.02°C.

When using our Vaisala Veriteq system, the design of the sensor impacts the calibration of the probes. The sensor has a stepped seal, so the neck of the probe is larger than the shaft of the probe and the measuring tip. This means that a standard drilled calibration bath, would create an air pocket around the sensor during the calibration. The ideal solution would be to use one of our fluids calibration baths, this however is not always practical when performing onsite calibrations. So we had an insert fabricated to ensure accurate calibrations using our Jofra RTC-159 Calibration Bath. The insert is designed to allow the sensing tip to be placed correctly.

Vaisala Veriteq Insert Design

As with any calibration, it is important to ensure that sensors are positioned correctly. Sensors must be fully inserted into the well as temperature varies throughout the axial gradient. The RTC-159 has a unique dual-zone heating control system, which keeps the sensors uniform throughout the block despite thermal load.

Our engineers take pride in every aspect of the qualification process to ensure the highest quality calibration and validation. Using the suitable baths an inserts is one of the many ways our engineers work to continuously improve quality.  

Luís Vaz

New Projects Validation Manager | Hikma Farmacêutica Portugal

5 年

Good article, totally agree, specially with the insert tube dimensions, we have been optimizing to assure that the dimensions of the probe fit and are not oversized inside the insert tube. Here we use Ametek/Jofra and Fluke calibrators.

Michael Aliberti

Metrology Manager at New England Biolabs

5 年

Great article and also a great topic, one that usually is overlooked in the industry. Our Validation team at Masy has the benefit of having a primary Metrology lab in-house. All of our equipment used by both the validation team and our field calibration team has all of the associated uncertainties quantified as part of the system. This, much like what you are pointing out in your article, ensures our teams are using instruments that have been scrutinized by a team of Metrologists, accredited through NVLAP. Great article, thanks for sharing!

要查看或添加评论,请登录

Chris Maughan的更多文章

社区洞察

其他会员也浏览了