Temperature influence in flow and volume measurement.

Temperature influence in flow and volume measurement.

Temperature influence in flow and volume measurement

In mechanical ventilators, the inspiratory and expiratory volume (Vti, Vte) and minute volume (Vi and Ve) must be precisely checked with a gas flow analyzers or a ventilator tester.

Ventilator testers, from different manufacturers, can measure flow and volume. However, flow accuracy is strongly temperature dependent and needs to be temperature compensated. This is particularly important because in turbine- operated ventilators raise the gas temperature above the ambient temperature.

No hay texto alternativo para esta imagen

A turbine driven homecare ventilator (Resmed Stellar shown above) raises the temperature of the respiratory gas to?30.2°C?due to the turbine.


Settings:

IPAP = 30cmH2O,

EPAP = 7cmH2O,

Rate = 15bpm.

The IMT Analytics?FlowAnalyser PRO?measures the flow with an accuracy of 1.65% of reading and the?CITREX H5?measures the flow with an accuracy of 1.90% of reading.

Some other gas flow analysers measures the flow with an accuracy of 1.70% of reading ONLY if gas temperature is between 19…27°C. That means at a gas temperature of 30°C, this gas flow analysers measures flow with an accuracy of 2.3% of reading (1.7%+3x0.2%) because per degree Celcius the flow accuracy must be corrected by 0.2% (see users manual).

No hay texto alternativo para esta imagen

Summary:

The?temperature influence?of flow and volume measurement on ventilator testers?must be taken into account. This is especially important when testing turbine-driven ventilators that deliver warm air. All IMT Analytics Gas Flow Analyzers feature a very complex patented?temperature compensation of the?flow measurement, which guarantees a constant flow accuracy over a wide temperature range.