A Step-By-Step Guide to "Verifying Analytical Valve Operation"
Emerson 700XA GC

A Step-By-Step Guide to "Verifying Analytical Valve Operation"

Ever wondered about verifying the operation of an analytical valve in a Gas Chromatograph ?

It's not as straightforward as observing the vent flow due to the continuous carrier gas flow.

Here's a quick and effective method to determine the functionality of your analytical valves while maintaining safety measures:

Power on GC, but ensure to turn off all Oven heater and other heat supplies.

Introduce the carrier gas (He or N2) at the sample inlet.

Forcefully activate the Sample Shut-Off Valve, if available.

Physically Open the sample outlet port of the analytical valve.

Carefully toggle the Analytical Valve ON and OFF via valve settings.

You will easily identify the switching of gas if the valve operates.

Remember, this is the last step after troubleshooting the solenoid valve and Actuation gas.

Vilas Baburao Sawant

Adage Automation Private Limited | VP & Head of Center of Excellence for Analyzers and Solutions

1 年

Confirming the vent flows is the most important step to start any GC which will automatically confirm the valve operation and leaks if any.

Dirk Horst

Freelance Trainer Process Analyzers & Sampling Systems and Custody Transfer / Fiscal Metering,

1 年

Only check of valve switching? Then also check for leakage in both positions and at operating temperature!

要查看或添加评论,请登录

社区洞察

其他会员也浏览了