Non-Recall of Accu-Chek Meters
Ron LaPedis
Cyber, BCP, and law enforcement evangelist & story teller. Ponemon Distinguished Fellow.
Do you use a Roche Accu-Chek Guide blood glucose meter with a serial number below 92311000001? The company issued a notice of "urgent medical device correction," which appears to be a warning rather than an actual recall. The letter states that Roche has reports of two serious injuries due to power problems with the meter, where it throws an error, doesn't power on, or batteries don't last very long.
While I received a letter from them outlining the problem, there is no information about this non-recall online, nor is there an offer to replace the defective units. Here is the text of the letter:
Issue: As part of our ongoing quality monitoring and marketing surveillance processes, Roche has identified that Accu-Chek Guide blood glucose meters may:
- Display E-9 Errors (indicating batteries need to be replaced
- Show unexpected low battery icons
- Have short battery life
- Not power on
We have thoroughly investigated this issue to identify root cause and have begun implementing appropriate corrective measures. This includes design improvements to make the meters more robust against battery performance issues.
Clinical Significance: Unexpected power issues with the Accu-Chek Guide blood glucose meters may result in temporary meter unavailability that could lead to a delay in treatment and/or therapy decisions. Such a delay may potentially lead to an adverse medical event or consequence. Serious injuries have occurred or could occur due to the failure mode associated with this recall. We have reports of two serious injuries.
Affected Product: Accu-Chek Guide blood glucose meters with a serial number below 92311000001. The serial number can be found on the back of the meter. Note: No other Accu-Chek products are affected.
Actions for users of Accu-Chek Guide blood glucose meters:
- Always have a spare set of batteries. The meter requires two (2) size CR2032 batteries.
- Be aware that battery life may vary due to factors such as temperature and battery manufacturer.
- Have a back-up testing method available.
- If you have questions regarding this notification or you need further support, please contact our Accu-Chek Customer Care team through our website (accu- chek.com), email address ([email protected]), or call (1-800-858- 8072).
- If there is a Business Reply Card enclosed with this mailing, please complete and return it as instructed on the Business Reply Card.
- Keep a copy of this notification for your records.
If the meter is powered on and displaying an E-9 error or low battery icon:
- Move the meter to a more moderate environment, such as ambient room temperature, to ensure optimal battery performance.
- Turn the meter off, then turn the meter back on.
- If the problem persists, perform a meter reset by doing the following: Remove the batteries and press and hold the Power/Set/OK button for at least 2 seconds then reinsert the batteries.
- If the problem still occurs after performing a meter reset, then replace the batteries with new ones according to the instructions in the user manual. Use high quality batteries in the meter, such as Panasonic. The meter requires two (2) size CR2032 batteries.
- If replacing the batteries does not correct the situation, please contact our Accu-Chek Customer Care team through our website (accu-chek.com), email address ([email protected]), or call (1-800-858-8072).
If the meter will not power on or has short battery life:
- Move the meter to a more moderate environment, such as ambient room temperature, to ensure optimal battery performance.
- Perform a meter reset by doing the following: Remove the batteries, press and hold the Power/Set/OK button for at least 2 seconds, then reinsert the batteries.
- If the problem still occurs after performing a meter reset, then replace the batteries with new ones according to the instructions in the user manual. Use high quality batteries in the meter, such as Panasonic. The meter requires two (2) size CR2032 batteries.
- If replacing the batteries does not correct the situation, please contact our Accu-Chek Customer Care team through our website (accu-chek.com), email address ([email protected]), or call (1-800-858-8072).