Meter Substitution: Is it Worth the Additional Headaches?
Morehouse Instrument Company
We create a safer world by helping companies improve their force and torque measurements.
We have been asked more frequently about the benefits of calibrating everything as a system versus calibrating the load cells and indicators separately and intermixing them using a calibrated load cell simulator.
Being able to intermix any load cell and meter combination makes much sense.
If someone needs two load cells and a meter to do a specific job, they can check out the two load cells and a meter and adapters and perform the calibration or test.
The other unused load cells would then be available for someone else. If done correctly, this would be quite beneficial.
However, many disadvantages and obstacles remain, such as maintaining(calibrating) a load cell simulator with enough span points, establishing metrological traceability, and correctly calculating its measurement uncertainty.
Every meter needs calibration, contributing to a larger overall measurement uncertainty.
As you do not have a married system, you need to establish traceability for both the load cell and meters independently.
Let's start by dealing with ISO 376 and ASTM E74 standards requirements.
These are standards required for calibrating force-proving instruments, most commonly known as load cells, to calibrate other force-measuring instruments, force machines, hardness machines, and testing machines, using ASTM E74, ASTM E4, ASTM E-10, ASTM E-18, ISO 376, ISO 7500 and so on.
ISO 376 and ASTM E74 requirements for meter calibration
ISO 376 in section C.2.11 Effect of a replacement indicator states,
?"The deviation between the two indicators should be determined (there are several methods, e.g., calibration of both indicators, use of a common bridge simulator), and the uncertainty of this deviation should be estimated (including factors such as the calibration uncertainty of the indicators and the stability of the common bridge simulator).
If corrections are made, the uncertainty of the deviation should be taken into account. If no corrections are made, the deviation and its uncertainty should be considered."
Section 5 of the standard goes into more detail about when an electrical measurement is made. It goes on to list conditions that must be fulfilled, such as:
Additionally, ISO 376 mentions programming indicators using span points.
If one does not use the calibration equation and programs point into an indicator that allows points from the calibration curve to be input so that the display is in units of force or torque but carries out linear interpolation between these points, the effect of this approximation to the curve should be investigated, and an uncertainty contribution should be included.
ISO 376 section 3.1 defines a force-proving instrument as a "whole assembly from the force transducer through to, and including, the indicator."
One might be thinking, I do not calibrate following ISO 376. One could only use the ASTM E74 standard or a commercial calibration.
ASTM E74 is very similar in the requirements for substitution.
Section 12 is explicitly titled Substitution of Electronic Indicating Instruments Used with Force-Measuring Systems.
The standard acknowledges that it might be desirable to treat the indicator and force-measuring instrument separately.
A huge benefit is that if you purchase the same indicators, one could be used as a backup if the primary unit fails.
The expensive calibration of the entire system could be avoided.
Then, the standard lists conditions that shall be satisfied to substitute a metrologically significant element of the electronic indicating instrument.
ASTM E74 Section 12.1.1 specifically states, "The electronic-indicating instrument used in the initial calibration and the instrument to be substituted shall each have been calibrated and their measurement uncertainties determined. The electronic indicating instrument to be substituted shall be calibrated with traceability to the SI over the full range of its intended use, including both positive and negative values if the system is used in tension and compression.
The calibrated range shall include a point less than or equal to the output of the force transducer at the lower force limit and a point equal to or greater than the output of the force transducer at the maximum applied force. A minimum of five points shall be taken within this range. The measurement uncertainty of each electronic indicating instrument shall be less than or equal to one-third of the uncertainty for the force-measuring system over the range from the lower force limit to the maximum force."
The section then states, "12.1.3 Excitation voltage amplitude, frequency, and waveform shall be maintained in the substitution within limits to ensure that the effect on the calibration is negligible. It is a user responsibility to determine limits on these parameters through measurement uncertainty analysis and appropriate tests to ensure that this requirement is met."
To summarize, you must calibrate a simulator to comply with the standard. The simulator must be capable of positive and negative values if the load cells are used in both positive and negative directions. The simulator must have at least one point less than or equal to the lowest force point value in the range and one for the highest point.
The first point is 0.5 mV/V, and the last one is 4 mV/V. If someone had a 4 mV/V (10,000 Force Units) load cell and the verified range of force was 500 through 10,000 Force Units, the simulator at 0.5 mV/V would be 1,250 Force Units. If the verified range of forces started at 200, a 0.08 mV/V first step would be required.
Note: The best high-end simulators typically have the first step of 0.04 mV/V or lower, as 0.04 mV/V on a 2 mV/V load cell equates to a 2 % llf. A simulator that starts at 0.1 mV/V would equate to a 5 % llf on a 2 mV/V load cell.? ?
2 mV/V is 5,000 Force Units. The end-user would need to raise their Class A verified range of forces to 1,250 FU using this simulator. This situation does not work for many, as they want to capture force values from the first non-zero calibrated point, typically below 5 % of the load cell's capacity.
The ASTM E 74 standard provides further guidance by stating that the indicator's measurement uncertainty shall be determined by one of the methods in Appendix X2. The appendix is non-mandatory, and there are likely other ways to establish a traceability path. However, section 12 is mandatory, and the end-user must abide by it.
In no case can someone substitute a meter without meeting the measurement uncertainty and traceability requirements.?
The standard further recommends that the simulator have a series of mV/V steps of the measurement range with similar impedance characteristics and states this requirement in section 12.1.2; thus, this criteria needs to be adhered to.
"The measurement uncertainty of the transducer simulator shall be less than or equal to one-tenth of the uncertainty for the force-measuring instrument." ASTM E74 further states, "Excitation voltage amplitude, frequency, and waveform shall be maintained in the substitution within limits to ensure that the effect on the calibration is negligible.
It is a user's responsibility to determine limits on these parameters through measurement uncertainty analysis and appropriate tests to ensure that this requirement is met. Substitution of an interconnect cable can have a significant effect on calibration. If an interconnect cable is to be substituted, see Note 15."
Appendix X2 details steps necessary to start to determine? the uncertainty and recommendations for calibrating the meter for substitution. Morehouse fully supports ASTM E74 and feels the membership is incredible. For around $100.00, one can join and get access to a catalog of standards. This author's opinion is that this is one of the best deals in the industry. Signing up is simply at astm.org. The E-28 committee is responsible for standards such as ASTM E74, ASTM E2428, ASTM E10, ASTM E18, and ASTM E4.
Summary of Top Requirements for Load Cell Simulator Calibration Needed for? Meter Substitution
The summation of what is needed is as follows:
Do the benefits outweigh the additional headache of using a simulator and being able to separate one's load cells from the indicator or decouple the system?
If you still think they might, there are many more error sources one needs to be aware of.
领英推荐
One of the error sources we see missed quite a bit is not having the load cell calibration excitation and waveform match that of the meter being used. A good example is many USB digital indicators providing less than 10 V DC output. To comply with these requirements, one must calibrate their load cells at the same excitation and waveform as the meter. If the USB meter puts out 5 VDC, the load cell calibration should be at 5 VDC.
Additional error sources include Calibration Uncertainty (Gain Error), Zero Offset, Temperature Effect on Sensitivity, Quantization Error, Normal Mode Voltage, Power Line Voltage Variation, Non-Linearity, Temperature Effect on Zero, Gain and Zero Stability, Common-Mode Voltage, Noise, Electrical Loading, Error Signals due to thermal EMF, Difference in cabling if not a true 6-wire system. All these error sources should be evaluated.
When it comes to calibrating load cells, it is important to comply with certain requirements in order to ensure accurate measurements. This means calibrating the load cells at the same excitation and waveform as the meter being used. For example, if a USB meter outputs 5 VDC, then the load cell calibration should also be done at 5 VDC.
However, other factors can contribute to errors in load cell measurements. These include Calibration Uncertainty (Gain Error), Zero Offset, Temperature Effect on Sensitivity, Quantization Error, Normal Mode Voltage, Power Line Voltage Variation, Non-Linearity, Temperature Effect on Zero, Gain and Zero Stability, Common-Mode Voltage, Noise, Electrical Loading, and even thermal EM.
When the amount of work required to capture all of these error sources, dot all of the I's and cross all of the T's is considered, we often find it much more sensible to consider buying one digital indicator per load cell. Most of these USB-type digital indicators, like the Morehouse HADI, are excellent, very reasonably priced, and can easily accommodate the goal of scheduling equipment.
If a load cell or a meter malfunctions, only one piece of equipment is down, and work can likely continue.
Morehouse can work with anyone to get that equipment back up quickly. If a HADI indicator gets run over, dropped, or damaged, we could quickly replace and calibrate the load cell with the new meter as a system—in less than two weeks.
Marrying one indicator to a load cell is often less risky and more cost-effective. It provides all the benefits of using different-size load cells for various measurements. Going down the substitution route requires separate calibration for the load cell and each indicator.
The long-term cost is typically much larger, as is the overall measurement uncertainty.
You are effectively now paying for two calibrations instead of one system calibration.
Typical Error Sources for Meter Substitution
When calculating Measurement Uncertainty for a meter to be used for substitution, the following are typical error sources:
Simulator Uncertainty includes the meter's resolution, simulator calibration and the associated reference standard uncertainties, the simulator's stability, and ratio uncertainty. At Morehouse, we achieve about ± 0.00005 mV/V uncertainty on our high-end simulator using different cables for positive and negative output, as the polarity switch introduces additional uncertainty.
On the meter side, Non-linearity, Stability, Environmental, Ref Uncertainty from the Simulator, Additional Cable Uncertainties, Noise or Resolution, Repeatability, and Reproducibility.
In our experience, most who use meter substitution add about 0.02 %—0.04 % uncertainty to their systems. This is too much uncertainty for ASTM Class AA calibrations, which are expected to be better than 0.05 %, too much for ISO 376 Class 00, 0.5, and likely too much for Class 1 & 2. The requirement for an ASTM Class A is to be better than 0.25 %. The contribution to uncertainty is often significant, though somewhat manageable.
I have discussed this topic with NIST and was told that they do not assign a Class A or AA-verified range of forces when calibrating load cells sent in without an indicator. This is an interesting debate for those calibrating following ASTM E74, as the LLF should be different when substituting meters, as the additional uncertainty would impact the verified range of forces.
Morehouse does offer calibration of load cell simulators to comply with either standard. Below is a page from our calibration report for one of our reference load cell simulators. The load cell simulator was explicitly designed to calibrate meters.??
These simulators have high-quality aged resistors and steps from 0.04 – 4.4 mV/V. The standard deviation is less than the resolution, hence the importance of having the resolution as part of the overall measurement uncertainty.
We have written a technical paper on the importance of instrument resolution in an uncertainty budget. That paper can be found here.
?
Morehouse Budget Load Cell Simulator
Back to the simulator, calibrating the meters for substitution is likely not good enough with this meter. Why would anyone want this? The best answer is cost. The simulator costs under $600.00 compared with a higher-end model costing over $4,500.00 plus calibration. So, what? It does not allow me to calibrate my meter. That is technically correct, though it is a very powerful tool.
Our simulator allows the end user to do the following:
Load cell Simulator - Conclusion
If you want to read more, please purchase the ISO 376 and ASTM E74 standards. If you are performing meter substitution, you should have controlled copies of any standard you try to comply with.
When accessing overall measurement uncertainty, ?I always strive to do what yields the lowest overall measurement uncertainty to limit the overall risk. Calibrating everything as a system is much better, keeping the measurement traceability chain clean.
Adding a load cell simulator and its associated measurement uncertainty and the calibration of several meters can be challenging.
Many people struggle with calculating measurement uncertainty for "married" systems, and adding more requirements creates additional uncertainty and headaches.
There is a risk/reward scenario for separately calibrating the indicator and load cells. Much additional work is required to comply with either ISO 376 or the ASTM E74 standard. It might be worth it if that extra work saves time and money.
Any meter used for substitution must have the same characteristics, excitation, and waveform.
Plus, the overall uncertainty increases by an additional 0.02 – 0.04 %, which will be absorbed by everyone else down the metrological traceability pyramid.
Though not suitable for meter calibration following ISO 376 or ASTM E74, our budget simulator can save much time when troubleshooting equipment and verifying that everything was keyed incorrectly via coefficients from a calibration report without breaking the bank.
We can provide higher-end simulators for indicator substitution, though the cost is likely over $5000.00, depending on the exact steps and requirements.
If you enjoyed this article, check out our LinkedIn and YouTube channel for more helpful posts and videos.
At Morehouse, we educate our customers to comply with the published requirements from ASTM E74 and ISO 376 regarding load cell simulator calibration and meter substitution.
We have been in business for over a century and focus on being the most recognized name in the force business. That vision comes from educating our customers on what matters most and having the proper discussions about force calibration concepts so everyone uses the proper methods.
We believe in changing how people think about force and torque calibration in everything we do. We challenge the "just calibrate it" mentality by educating our customers on what matters and what causes significant errors and focusing on reducing them.
In addition, Morehouse makes simple-to-use calibration products. We build excellent force equipment that is plumb, level, square, and rigid. We provide unparalleled calibration service with less than two-week lead times.
Contact us at 717-843-0081 to speak to a live person or email [email protected] for more information.