Making Sense of SF6 Analyzer Results: A Quick Guide
Interpreting the results of an SF6 analyzer measurement can be a bit daunting, especially if you are not familiar with the various parameters that are measured and the units in which they are reported. In this blog post, we will discuss some of the key parameters measured by an SF6 analyzer and how to interpret the results.
The 3 Standard Gases
Understanding Sensor Tolerances
The accuracy of an SF6 analyzer’s measurements is affected by the tolerances of the individual sensors used to measure each parameter. Sensor tolerances refer to the degree of accuracy and precision that can be expected from the sensor in ideal laboratory conditions. In the field, conditions may be less than ideal, which can further affect the accuracy of the measurement.
Here’s an example: The SF6 purity sensor might show a “97%” reading. The actual SF6 purity value may be as low as 95% or as high as 99%, given the tolerance range. Therefore, it is important to consider the acceptable range and the potential sources of error when interpreting the results of an SF6 purity test.
If you are experiencing some weird test results, try taking a few tests and averaging them to see where you fall. If you are having massive swings in measured values that are beyond what you expected given the sensor tolerances, try flushing your system with a bottle of nitrogen. If the device is still operating abnormally after that, it might need to be recalibrated or repaired.
Dew Point vs. ppmv for Moisture
When it comes to measuring moisture in a gas, two common methods are used: parts per million by volume (ppmv) and dew point temperature. Both methods are used to express the amount of moisture in a gas, but they use different units and have different relationships.
领英推荐
PPMV is a unit of measurement that expresses the concentration of a gas in a mixture of gases. In the case of measuring moisture, ppmv expresses the amount of water vapor in a gas relative to the total amount of gases present in the mixture. For example, if a gas mixture contains 1 ppmv of water vapor, that means that for every 1 million molecules of gas in the mixture, 1 molecule is water vapor.
Dew point, on the other hand, is a measure of the temperature at which the water vapor in a gas will begin to condense into liquid water. It is expressed in degrees Celsius or Fahrenheit. When the temperature of a gas falls below the dew point, the water vapor will start to condense out of the gas and form liquid droplets.
The relationship between ppmv and dew point is influenced by factors such as temperature and pressure. Generally, as the dew point of a gas decreases, the ppmv of water vapor in the gas also decreases. This is because as the gas cools, it becomes less able to hold moisture, so the water vapor condenses out of the gas and the concentration of water vapor in the gas decreases.
Conversely, as the dew point of a gas increases, the ppmv of water vapor in the gas also increases. This is because as the gas gets warmer, it becomes more able to hold moisture, so the concentration of water vapor in the gas increases.
Therefore, to accurately measure the amount of moisture in a gas, it’s useful to consider both the ppmv of water vapor and the dew point temperature. By measuring both of these parameters, it is possible to obtain a more complete picture of the moisture content of the gas. Handy dew point calculator –?https://dew-point.com/dew-point_calculator.asp
Verify Sensors are Calibrated & Accurate
To check the accuracy of an SF6 analyzer’s sensors, you can use calibrated gases with known certified gas concentrations. Here are the steps to follow: