HPLC Insights : Negative Peaks and Baseline Drift
Mohammed Ksassoua
Senior Lab Chemist | QC Raw Materials Specialist |12Y+ of Expertise in QC |16K+ Let's Connect ????
How to Address Negative Peaks and Baseline Drift in HPLC?
Overview
Negative peaks and baseline drift are relatively uncommon in HPLC with UV detection. However, they can arise under specific circumstances. This article will explore their causes, solutions, and practical considerations for troubleshooting these issues.
Example Figure for Negative Peak:
Negative Peaks: Causes and Solutions
Contrary to some troubleshooting guides, negative peaks are rarely caused by your analyte absorbing less than the mobile phase. Here’s why and how to resolve the issue:
1. Reference Wavelength Subtraction
Why It Happens:
In HPLC systems with diode array detectors (DAD), a reference wavelength is often set by default (commonly 360 nm with a bandwidth of ±50 or ±100). This reference wavelength subtracts noise (e.g., lamp noise or refractive index effects) from the signal at your sample wavelength (e.g., 254 nm).
If your compound absorbs significantly at the reference wavelength, the subtraction may yield a negative peak.
Solution:
Turn Off the Reference Wavelength:
If your compounds are highly colored (e.g., red, orange, green, or purple) and absorb in the visible range, disable the reference wavelength. This will eliminate negative peaks without significant loss of sensitivity.
Adjust the Reference Wavelength:
Move the reference wavelength to a region
where your compound does not absorb.
2. Considerations for Detection Wavelengths:
Most compounds absorb strongly at 254 nm (e.g., benzene rings) or 220 nm (acid groups). Ensure that the detection and reference wavelengths are optimally configured for your analytes.
3. Mobile phase absorption/refraction tooh high
Use less absorptive/refractive solvents
or mobile phase additives.
Change UV wavelength or switch the
detector polarity.
4. Injection solvent effects
If possible, use the starting mobile phase composition or weaker solvents as the injection solvent.
Ensure that the injection solvent is miscible with the mobile phase and causes no salt precipitation.
5. Improper method settings
Check the detector settings, such as polarity settings, and optimize them if needed.
Baseline Drift: Causes and Solutions
Baseline drift is typically minimal in HPLC unless you’re working with refractive index (RI) detectors or conducting gradient runs.
Example Figure for baseline Drift:
1. Refractive Index Effects in Gradient Runs:
Why It Happens:
Baseline drift can occur due to refractive index differences between the initial and final mobile phases during a gradient run.
Solution:
Turn on the reference wavelength. The subtraction of noise across all wavelengths stabilizes the baseline. The default setting of 360 nm with a bandwidth of ±100 is sufficient in most cases.
2. Absorbance of Mobile Phase
Negative drift
Use non-absorbing mobile phase solvents and modifiers.
Use high-purity HPLC grade or better solvents only.
Positive drift
Use higher UV-absorbance detector wavelengths where analytes are still present.
Use non-absorbing mobile phase solvents and modifiers.
3. Room Temperature Fluctuation
Ensure the system is positioned in a laboratory location free of temperature fluctuations.
Use a column thermostat to control column temperature.
4. Mobile Phase Miscibility
Ensure that the mobile phases are completely miscible.
领英推荐
5. Replace the column if necessary
The eluent pH and the column temperature should never exceed the manufacturer‘s specifications. Use specialty columns when working at high temperatures and/or pH extremes.
Most silica-based columns degrade rapidly at pH below 2 or above 8, especially at temperatures higher than 40 °C.
6. Detector Cell Contamination
Inspect the flow cell for salt crystals, air bubbles, or contamination. If necessary, clean/flush the flow cell or replace it.
7. Aging Detector Lamp
Check if the on-time hours of the detector lamp have exceeded the manufacturer’s specifications.
Run the detector lamp intensity test or equivalent to check the condition of the lamp. Replace the lamp if necessary.
Noisy Baseline
1. Air Bubble in the System
Degas mobile phases.
Purge the pump and system with a viscous mobile phase to remove any air bubbles.
Due to its miscibility with water and most organic solvents, isopropanol is often used to remove air bubbles from an HPLC system.
2. Improper Method Settings
Check the detector settings, such as wavelength and data rate, and optimize them if needed.
3. Aging Detector Lamp
Check if the on-time hours of the detector lamp have exceeded the manufacturer’s specifications.
Run the detector lamp intensity test or equivalent to check the condition of the lamp. Replace the lamp if necessary.
4. Flow Rate Inconsistencies
Ensure that the method settings are properly set and the gradient system is delivering the correct gradient composition.
Check if the mobile phase reservoirs are empty or if the filter frits are blocked.
Check for air bubbles in the system. If necessary, purge the pump and system with a viscous mobile phase to remove any air bubbles.
Check the system for any leaks.
Check for leaking pump seals. If necessary, replace pump seals.
Check for sticking check valves. If necessary, sonicate or replace check valves.
Note:
The eluents should be miscible and should not form salt precipitates over the range of the entire gradient. The eluents should also be properly degassed.
Due to its miscibility with water and most organic solvents, isopropanol is often used to remove air bubbles from an HPLC system.
5. Detector Cell Contamination
Inspect the flow cell for salt crystals, air bubbles, or contamination. If necessary, clean/flush the flow cell or replace it.
6. Improper Mixing of Mobile Phases
Ensure mobile phase additives, such as TFA, are homogeneously mixed before reaching the detector flow cell.
Note:
Consider changing the mobile phase mixer in the system flow path to generate more homogeneous mobile phases and more consistent gradient formations.
7. Mobile Phase Miscibility
Ensure that the mobile phases are completely miscible.
8. Electrical Interference
Ensure that the power supply to the LC system is stable.
Check for local interfering sources.
If possible, use an independent electrical current for the LC system.
Practical Review: Negative Peaks and Baseline Drift
1. Negative Peaks:
If observed, disable the reference wavelength or adjust it to a non-absorbing region for your analytes.
Colored compounds are often the culprits due to their absorption in the visible spectrum.
2. Baseline Drift:
Ensure the reference wavelength is activated to counter refractive index effects during gradient runs.
3. Sensitivity vs. Practicality:
While using a reference wavelength enhances sensitivity (e.g., reducing detection limits from 1 ppm to 0.7 ppm), it’s not critical for routine analyses. Prioritize eliminating artifacts such as negative peaks when necessary.
Final Thoughts
Understanding and addressing negative peaks and baseline drift involves a balance of technical adjustments and practical decision-making. Whether you’re optimizing your diode array detector settings or refining your gradient methods, these insights can help you troubleshoot effectively.
Stay connected for more updates and expert tips on HPLC and chromatography techniques..
For more insights into chromatography, join the Chromatography Knowledge Exchange Group ??
https://www.dhirubhai.net/groups/12954239
Analytical Research Chemist ?? #P?EP-HPLC#HPLC.#.#Flash chromatography.(Biotage or Grace)# Greater Hydrabad Area. Genome valley.(IKP PARK)
1 个月It's very informative
Working in a Technology Transfer and core validation team
1 个月Very informative