Determination of Furan in Coffee by Static and Dynamic Headspace/GC/MS

Determination of Furan in Coffee by Static and Dynamic Headspace/GC/MS

Introduction

Coffees have aroma and taste based on where the beans are grown, roasting techniques and flavour additives.?One component of the bean responsible for the aroma and some of the taste are volatile organic compounds (VOC).?Furan is one VOC that has recently been investigated in food.

Figure 1 shows the SCION Instruments 8X00 GC platform.

Geen alternatieve tekst opgegeven voor deze afbeelding

Figure 1. the SCION Instruments 8500 GC platform?with 8700 SQ GCMS.?

An FDA method for furan in food uses static headspace and requires 30 minutes of incubation time at?60°C.?

This application note presents data comparing the determination of furan in dry coffee by both static and?dynamic headspace techniques. In the dynamic headspace method analytes are swept and concentrated on a sorbent trap that greatly increases compound sensitivity and decreases sample preparation time.

Table 1 is showing the analytical conditions of the GC-MS system.

Geen alternatieve tekst opgegeven voor deze afbeelding

Table 1. the SCION Instruments 8500 GC platform?with MS TQ

Instruments Parameters

The HT3 headspace sampler was connected to a SCION Instruments 8500-GC-MS. In figure 2 the SCION Instruments HT3 Headspace sampler is presented.

Geen alternatieve tekst opgegeven voor deze afbeelding

Figure 2. SCION Instruments HT3 Headspace Sampler.

?The Headspace sampler?parameters are presented in table 2.

Geen alternatieve tekst opgegeven voor deze afbeelding

Table 2: Parameters HT3 Headspace sampler.

?Experimental

Coffee samples of 5 grams were weighed into vials and crimp capped.?The static method followed the FDA’s 30 minutes sample heating time.?The dynamic method trapped the VOC’s on a proprietary #9 trap by sweeping the sample for 15 minutes at 50mL/min. The VOC’s were rapidly desorbed from the trap at 250°C and detected with the GC/MS.?The HT3 maintained the sample at 60°C following the FDA method.

Results and Discussion

A sample chromatogram is presented in figure 3, showing the chromatogram The peak area of the 68m/z ion specific for furan with both the static and the dynamic methods were compared.?Table 3 is the peak area data and its corresponding graph.?

Geen alternatieve tekst opgegeven voor deze afbeelding

Figure 3:?Comparison of the Dynamic (Blue) and Static (Red) Total Ion Chromatogram (TIC) (top) and the Specific Mass Ion 68 m/z for Furan (bottom).

The dynamic technique detects approximately 30 times more furan from the same size sample when compared to a static method. In addition the dynamic option shortens the analysis time to 15 minutes.

Geen alternatieve tekst opgegeven voor deze afbeelding

Table 3: Precision of the quantitation column.

Conclusion

The determination of furan in coffee in association with the SCION Instruments 8500 GC-MS is easy to perform both static and dynamic techniques with the?HT3 SCION Instruments Headspace Sampler.

Although the 4X6-GC series is not shown in this application note, it is also possible to perform this method on the SCION instruments 4X6 GC series. Ordering information can be found below. For customisation please contact your local sales representative.

Order Information

Geen alternatieve tekst opgegeven voor deze afbeelding

Download Application Note

Download complete Application Note here:?Determination of Furan in Coffee by Static and Dynamic Headspace/GC/MS

Stéphan Laperrière, M.Sc.

I Can Help | Positive Problem-Solver | Scientific Approach | Eternal Student-Athlete | Kids First | Ex-Pro Hockey Player

2 年

Thanks for the post SCION Instruments. Very nicely illustrated, I was unaware of such a potential difference. I am not much of a coffee guy, but what is the range of roasting temperature compared to the ones seen in the analysis process (like the step of "desorbed from the trap at 250°C?"). Is the term "Specific mass ion" something within Scion, or that particular system type? Just curious because I have been more exposed to "Extracted-ion chromatogram" (XIC, EIC). As an aside, it is unfortunate that the "native" axis and linked infos do not come out clearly. The bold overlay within the graphs is well done and makes it really clear. I would consider putting both chromatograms on the same scale.