Cleaning and Maintenance of HPLC Pump
Jairus Fernandes
Developing, Validating and Transferring of Analytical Methods for the Success of New Products ??
Cleaning an HPLC (High-Performance Liquid Chromatography) pump is a crucial step in maintaining the performance and accuracy of the system. Here's a general guide on how to clean an HPLC pump:
?
Before starting any cleaning procedure, consult the instrument manual provided by the manufacturer.
?
Prepare the Solvent:
Use high-purity solvents, such as HPLC-grade water, methanol, acetonitrile, or a mixture of these solvents.
Ensure that the solvents are filtered through a 0.2-micron filter to remove particulate matter.
?
Flushing the Pump:
Flush the pump with the solvent to remove any residual mobile phase.
Set the pump to deliver a flow rate higher than the typical operational flow rate.
?
Disconnect Tubing:
Disconnect the inlet and outlet tubing from the pump. Be cautious and follow the system's guidelines to prevent any spills.
?
Clean the Pump Head:
Remove the pump head according to the manufacturer's instructions.
Clean the pump head components with the solvent. Use lint-free wipes or swabs to gently wipe away any deposits or contaminants.
?
领英推荐
Clean the Inlet and Outlet Check Valves:
If applicable, remove and clean the inlet and outlet check valves. Soak them in the solvent and use a small brush to remove any debris.
?
Flush the Pump with Solvent:
Reconnect the tubing and flush the pump with the solvent for an extended period. This helps to remove any remaining contaminants from the pump internals.
?
Perform a Seal Wash:
Some HPLC systems have a seal wash feature. If your system has this capability, activate the seal wash to clean and flush the seals.
?
Inspect Seals and Tubing:
Inspect the pump seals and tubing for signs of wear or damage. Replace any worn or damaged components.
?
Perform a System Check:
Reconnect all tubing and perform a system check to ensure that the pump is operating correctly. Check for leaks and verify the flow rate accuracy.
?
Replace the Mobile Phase:
Replace the mobile phase with fresh, filtered solvent.
?
Run a System Suitability Test:
Run a system suitability test to ensure that the HPLC system is performing within acceptable parameters.
?
Remember to follow the specific guidelines provided by the instrument manufacturer. Regular maintenance and cleaning are essential for ensuring the reliability and accuracy of HPLC systems.
EL JADIDA MOROCCO
9 个月can you sent me this document
IT Certification at TIBCO
10 个月Transform your F5 Certification dreams into reality with www.certfun.com/f5. ?? Dive deep, practice hard, and achieve greatness! #CertFun #F5Dreams #TechJourney ??
Enabling Career and Business Growth through proven LinkedIn Strategies and Diplomatic Networking.
11 个月Very useful Jairus Fernandes I remember the old days with the Varian single piston pump ??
Pharmaceutical Specialist and Knowledge kit for Pharma industry
11 个月https://flairpharma.com/hplc-principle/
Quality Control Analyst
11 个月Thanks for sharing