Best practice for separation columns in ion chromatography (IC) – Part 2
The second part of this blog series about best practices for IC separation columns focuses on application-related topics that have an impact on the column suitability and stability. First, there is the proper choice of the column that best suits the intended application. Then we turn to the operating parameters which can be modified in order to optimize the separation between analytes, and what the respective effects and possibilities are.
Choice of column length and diameter
Metrohm offers a?broad range of columns?that contain different stationary phases, have different lengths and/or inner diameters. The choice of the stationary phase has a great impact on the selectivity between the individual analytes on the one hand, as well as the stability against different sample matrices on the other hand. Instead, the column length has no impact on the selectivity, but rather on the separation efficiency between the individual peaks.
Find out more about Metrohm’s wide selection of separation columns for ion chromatography in our?Column Catalog.
Effects of column length
Figure 1. Effect of column length on the retention times of the standard anions on the Metrosep A Supp 17 column (1: fluoride, 2: chloride, 3: nitrite, 4: bromide, 5: nitrate, 6: sulfate, 7: phosphate). Click image to enlarge.
In the following chromatograms (Figure 1), the effect of the column length on the separation efficiency for the?Metrosep A Supp 17 column series?is shown. Whenever choosing a column length, you should take the?complexity of the intended separation?and the presence of?matrix components?that could disturb the ions of interest into account.
Effects of column diameter
In addition to providing different lengths of IC separation columns, Metrohm also offers most columns in both in?4 mm?inner diameter and?2 mm?inner diameter (known as ?microbore?) versions. With regard to this, there are several criteria to distinguish:
Find out more about MARGA and its capabilities for continuous air quality monitoring in our?blog post.
Optimizing the analyte separation
Next to the column itself, several other parameters can be modified to optimize the?selectivity?of the separation. These parameters include temperature, eluent components and strength, and organic modifiers.Figure 2. Effect of temperature variation on the retention times of a suite of standard anions on the Metrosep A Supp 17 column (1: fluoride, 2: chloride, 3: nitrite, 4: bromide, 5: nitrate, 6: sulfate, 7: phosphate). Click image to enlarge.
Effects of modifying the temperature
One of the simplest ways to fine-tune the separation selectivity in IC is by modifying the temperature of the analysis. This is accomplished by using the integrated column oven in the instrument (if available). Multiple effects can be observed, for instance in anion analysis. As an example, the impact of the temperature on the selectivity is shown in the chromatogram overlay (Figure 2) for the?Metrosep A Supp 17 column line.?
Figure 2. Effect of temperature variation on the retention times of a suite of standard anions on the Metrosep A Supp 17 column (1: fluoride, 2: chloride, 3: nitrite, 4: bromide, 5: nitrate, 6: sulfate, 7: phosphate). Click image to enlarge.
Effects of modifying the eluent composition and strength
Eluent composition and strength can be used to change the elution order of several analytes while using the same separation column. In cation chromatography, a retention model was developed by P.R. Haddad and P.E. Jackson, which allows researchers to predict retention times when changing the eluent composition [1].
领英推荐
Considering that the column remains identical in each determination, no change of ion exchange equilibrium and column capacity is to be expected. Therefore, when changing?only the eluent concentration, the following correlation can be used:
If nitric acid is used as the eluent,?y?= 1, and the model can be simplified to:?
Where:
Applying this formula to practical situations in the laboratory means the following: with increasing the eluent strength, alkaline earth metals are accelerated much faster (x?= 2) in comparison with alkali metals (x?= 1), and thus it is possible to elute magnesium before potassium. This effect is called?electroselectivity.
Multivalent metal ions are capable of forming complexes with dedicated complexing agents. Therefore, selectivities can be modified by adding complexing agents to the eluent. As an example, dipicolinic acid (DPA) is often used to complex calcium, which leads to a reduction of the effective charge of calcium. As a consequence, the retention time of calcium is reduced and calcium elutes before magnesium in the chromatogram (Figure 3).
The retention of monovalent cations can be influenced by the addition of crown ether to the mobile phase.
Figure 3.?Effect of DPA concentration in the eluent on the retention times of several cations measured using the?Metrosep C 6 column.?
Anion systems are more complex regarding the retention time model, although the same electroselectivity effect can be observed to some extent for anions. However, when changing the eluent strength, the eluent pH also frequently changes, leading to different deprotonation equilibria of multivalent anions (e.g. phosphate). This influences the effective charge of the analyte, and by doing so, also influences its retention in a similar way as previously described for the effects of changing temperature.
In some cases, the use of a small amount of an organic modifier such as methanol, acetonitrile, or acetone in the eluent can make sense:
Learn more about the?Metrohm Inline Sample Preparation (MISP) techniques?offered here!
Be aware that the addition of organic modifiers will also affect the separation selectivities. For the standard anions, the effect is similar to that observed with increased temperatures: the peak shapes of the polarizable ions such as nitrite, bromide, and nitrate are improved.
Organic acids on the other hand may react very differently compared to the standard anions, and their reaction also strongly depends on the type of organic modifier used. Sample chromatograms that show the effect of the organic modifier on retention of analytes are shown in the manual for the?Metrosep A Supp 10?column.
Download the?Metrosep A Supp 10 column manual?here to see example chromatograms showing the effects of organic modifiers on analyte retention time.
For more information about column care, check out our?blog post for different tips and tricks.