5 Most Recommended Dental Handpieces - and How to Sterilize Them Properly According to their IFU
Reuben Druckenmiller
Helping to Keep Your Sterilizers Running With No Downtime
I recently came across an article in Dental Country written by Sean Myran, DDS, in which he lists his top 5 recommended dental handpieces. This got me thinking, what is the required sterilization parameters for each of these handpieces.
I decided to do the hard work for all of you and track down the IFUs for each of the 5 listed handpieces. Here are the 5 that Dr. Myran recommends:
1.??????MASTERtorque LX M8900 L by KaVo Dental
2.??????Titan 3 Low-Speed Handpiece by DentalEZ
3.??????Midwest Stylus Plus by Dentsply Sirona
4.??????INTRAflex LUX 2 2307 LN by KaVo Dental
5.??????430 SW Torque High-Speed Handpieces by DentalEZ
Now regardless of whether or not you agree with his list (Dental connections: Please feel free to comment on which handpiece you would recommend using and I will find their IFUs for you as well if you are interested) of the top 5 handpieces, let's take a look at each of their validated sterilization cycles.
1.??????MASTERtorque
a.??????Gravity Type Autoclave
????????????i.?????At least 10 minutes at 132C (270F) with a 30-minute dry cycle
b.??????Autoclave with 3-time pre-vacuum
????????????i.?????At least 3 minutes with a 16-minute dry cycle
2.??????Titan 3 Low-Speed
a.??????Gravity Type Autoclave
?????????????i.?????Minimum 15 minutes at 132C (270F)
????????????ii.?????Minimum 30 minutes at 121C (250F)
b.??????Pre-vacuum
?????????????i.?????Minimum of 4 minutes at 132C
3.??????Midwest Stylus
a.??????Gravity Steam Sterilization
????????????i.?????Full Cycle 121C (250F) for 30 minutes
???????????ii.?????Full Cycle 132C (270F) for 15 minutes
b.??????Dynamic Air Removal
?????????????i.?????Full Cycle 132C (270F) for 4 minutes
????????????ii.?????Full Cycle 135C (275F) for 3 minutes
领英推荐
4.??????INTRAflex Lux 2
a.??????Gravity Type Autoclave
?????????????i.?????At least 10 minutes at 134C (273F)
b.??????Autoclave with 3 times pre-vacuum
??????? ????i.?????At least 3 minutes at 134C (273F)
5.??????430 SW Torque High-Speed
a.??????Gravity Type Autoclave
?????????????i.?????Minimum of 15 minutes at 132C (270F)
?????????????ii.?????Minimum of 30 minutes at 121C (250F)
b.??????Pre-Vacuum Autoclave
???????????? i.?????Minimum of 4 Minutes at 132C (270F)
c.??????NOTE: Do not exceed 135C (275F)
Now you will begin to notice a couple of things if you read each of those sterilization cycles and you know a thing or two about how autoclaves work and what a gravity-type autoclave is versus an SFPP autoclave and again versus a Pre-Vacuum autoclave.
1.??????Most of them list a gravity and Pre-Vacuum, they do not include a validated sterilization exposure time for SFPP autoclaves (Midmark M9 and M11 new series). Leaves a big question mark for Midmark users.
2.??????If you do not have a Pre-vacuum autoclave (Something like Tuttnauer’s T-Edge Series, W&H’s Lexa, Statim’s Bravo, or Tuttnauer’s Elara 11) then you are looking at the gravity cycles - they are significantly longer exposure times.
3.??????Some handpieces, like the Midwest Stylus, can be sterilized with DYNAMIC AIR REMOVAL as opposed to directly calling out Pre-vacuum. While this is a hotly debated topic amongst those who understand the major differences in how desktop vs freestanding autoclaves work, technically a Midmark M9 or M11 is considered a ‘Dynamic Air Removal’ type autoclave as it uses what is called an SFPP (steam flush pressure pulse) and so, therefore, can be used to sterilize this handpiece. Also, as a quick side note, Dynamic Air removal includes two types of autoclave air removal methods – SFPP and Pre-Vacuum (Of Which Pre-Vacuum is the more effective method when it comes to desktop autoclaves)
Each of the IFUs can be found at the links below:
MASTERtorque – Click Here
Titan 3 -?Click Here
Midwest Stylus Plus - Click Here
INTRAFlex Lux 2 - Click Here
430 SW Torque High Speed - Click Here
If you want to read Dr. Myran’s article you can also find that here - Click Here
Weapon of Mass Microbial Destruction ? Serial Disruptor within Clinical Education ? Content Wizard ? Consultant ? Media ? Marketing ? Mayhem ? Clinical Sterile Processing Expert ? Catalyst (Networking) Converter
2 年Love this kind of educational content, Reuben Druckenmiller! This is the way ??
Helping to Keep Your Sterilizers Running With No Downtime
2 年As always Jim Scott - love to get your expert input here